Comparison of reduction rate of bacterial contamination of dental office air during dental procedures using conventional suction-high vacuum suction-aerosol suction device-betadine 1%-retractor
Abstract
Introduction:
Aerosols are one of the key means infections can spread in dental practice. It is important to take necessary measures in reducing aerosols. The purpose of this study is to compare bacterial presence in air using conventional saliva ejector, high vacuum suction, betadine 1% in reservoir and retractor and aerosol suction device.
Materials and Methods:
In this study a private office located in Tabriz which had the mentioned equipment was chosen. The study was conducted in 8 separate groups, single system or combined methods of saliva ejector, high vacuum suction, office fan, retractor, betadine 1% in reservoir and aerosol suction device. For each system 3 patients with class II restoration were examined. Samples were collected during the treatment from 6 different points at the same height as the patient’s head. Before each treatment the office was disinfected by UV light and the patients rinsed their mouth with betadine 1% mouthwash. betadine 1% was made by adding 1-gram betadine to 100mls distilled water and was added to dental unit reservoir. Using two suctions simultaneously one high vacuum suction was placed over retractor by the assistant and the conventional suction was placed inside patient’s mouth under retractor.
Results:
High vacuum suction had a statically significant effect in aerosol reduction(p<0.001). Using retractor was not statically significant. Adding office fan to all of the groups reduced bacterial colonies significantly(p<0.001). Adding betadine 1% to the reservoir had the strongest effect on bacterial colonies, 0.5-meter (46.58%) in 1.5 meters (50.68%), 2 meters (53%) and in pc (61.84%) reduction compared with high vacuum suction(p<0.001).
Conclusion:.
The best way to reduce aerosols in dental office is to use high vacuum suction, betadine 1% in reservoir with an office fan.
Collections
Related items
Showing items related by title, author, creator and subject.
-
Comparison of reduction rate of bacterial contamination of dental office air during dental procedures using conventional suction –high vacuum suction-aerosol suction device – ozone water – retractor
Navaie, Nima (Tabriz University of Medical Sciences, School of Dentistry, 2021)Introduction: Due to the type of dental treatment and the proximity of staff to patients, dental services can be a suitable environment for aerosol release. The use of aerosol reduction systems and methods is of great ... -
Neonatal Intensive Care Units Nurses' Attitude Toward Advantages and Disadvantages of Open vs Closed Endotracheal Suction.
Valizadeh, L; Janani, R; Janani, L; Galechi, F (2014)The vital issue of protecting the airway and maintaining ventilation in preterm infants makes tracheal suctioning an important procedure. The decision to use closed or open endotracheal suction method depends on the clinical ... -
Ventilator-associated pneumonia in patients admitted to intensive care units, using open or closed endotracheal suctioning.
Hamishekar, H; Shadvar, K; Taghizadeh, M; Golzari, SE; Mojtahedzadeh, M; Soleimanpour, H; Mahmoodpoor, A (2014)Critically ill patients under mechanical ventilation require frequent suctioning of airway secretion. Closed suction permits suctioning without disconnection from ventilator; so it might decrease hypoxemia and infection ...