COMPARISON OF RADIATION DOSE TO PATIENT AND STAFF FOR TWO INTERVENTIONAL CARDIOLOGY UNITS: A PHANTOM STUDY

Thumbnail
تاریخ
2008
نویسنده
Mesbahi, A
Mehnati, P
Keshtkar, A
Aslanabadi, N
Metadata
نمایش پرونده کامل آیتم
چکیده
The purpose of this investigation was to measure the patient and staff dose during routine interventional cardiology procedures for an image intensifier-based and a. at detector system using a water phantom. The Integris BH3000 image intensifier-based (Philips) and the Axiom Artis. at detector-based (Siemens) angiography units were used in this study. The accuracy of tubes potential and irradiation timers and also internal dosimeters were verified and confirmed. A water phantom with a thickness of 18 cm was used for patient and staff dose measurements. For the Philips system, phantom entrance dose rates were 2.77 and 38.97 mu Gym(2) s(-1) during fluoroscopy and cineangiography. The respective dose rates for the Siemens were 1.98 and 13.46 mu Gym(2) s(-1). Phantom entrance dose rate was 28.5 and 65% higher for the Philips system during fluoroscopy and cineangiography, respectively. Comparing the scattered dose rates at the operator location showed that the. at detector-based Siemens system delivers five times lower dose to the operator in comparison with the image intensifier-based Philips unit. The results suggest that the decrease in received dose of the patient and staff is achievable using the. at detector system. In addition, application of lead curtain and glass is recommended to lower the cardiologist dose especially for the image intensifier-based Philips system.
URI
http://dspace.tbzmed.ac.ir:8080/xmlui/handle/123456789/51266
Collections
  • Published Articles

مخزن دانش دانشگاه علوم پزشکی تبریز در نرم افزار دی اسپیس، کپی رایت 2018 ©  
تماس با ما | Send Feedback
Theme by 
Atmire NV
 

 

مرور

همه مخزنجامعه ها و مجموعه هابراساس تاریخ انتشارنویسنده هاعنوانهاموضوعاین مجموعهبراساس تاریخ انتشارنویسنده هاعنوانهاموضوع

حساب من

ورودثبت نام

مخزن دانش دانشگاه علوم پزشکی تبریز در نرم افزار دی اسپیس، کپی رایت 2018 ©  
تماس با ما | Send Feedback
Theme by 
Atmire NV