Patient's radiation exposure in coronary angiography and angioplasty: The impact of different projections
View/ Open
Date
2014Author
Farajollahi, A
Rahimi, A
Shal, EK
Ghaffari, S
Ghojazadeh, M
Tajlil, A
Aslanabadi, N
Metadata
Show full item recordAbstract
Introduction: We aimed to determine angiography projections with lower Dose Area Product (DAP) rate by measuring the mean DAP and fluoroscopy times in coronary angiography (CAG) and percutaneous coronary intervention (PCI) and calculating DAP rate in different projections. Methods: DAP and fluoroscopy times were measured in all employed projections in real-time in 75 patients who underwent CAG or PCI by a single cardiologist in Madani Cardiovascular University Hospital (45 in CAG group and 30 in PCI group). DAP rate was calculated in both groups and in all projections. The projections with highest and lowest DAP rate were determined. Results: Mean DAP was 436.73 315.85 dGy cm2 in CAG group and 643.26 359.58 dGy cm2 in PCI group. The projection 40 LAO/0 had the highest DAP rate in CAG group (28.98 dGy cm2/sec) and it was highest in 20 RAO/30 CR in PCI group (29.83 dGy cm2/sec). The latter projection was also the most employed projection in PCI group. Conclusion: The amount of radiation dose in this study is in consistent with the previous reports. Specific angiographic projections expose patients to significantly higher radiation and they should be avoided and replaced by less irradiating projections whenever possible. 2014 The Authors.