dc.description.abstractIntroduction: In radiation therapy centers across Iran, protection of normal tissues is usually accomplished by either Cerrobend or lead block shielding. In this study, the influence of these two shielding methods on central axis dose distribution of photon beam a Cobalt unit was investigated in clinical conditions. Materials and Methods: All measurements were performed for 60Co ?-ray beams and the Cerrobend blocks were fabricated by commercial Cerrobend materials. Standard lead block shields belonged to Cobalt unit. Data was collected through a calibrated ionization chamber, relative dosimetry systems and a TLD dosimetery. Results: Results of the percent depth dose (PDD) measurements at depths of 0.5, 1, 5, 10, 15 and 20 cm for 23 different field sizes of patients with head and neck cancer showed no significant differences between lead and Cerrobend shielding methods. Measurement results of absolute dosimetry in depths of 1.5, 3, 5, 7, 10 and 12 cm also showed no significant differences between these two shielding methods. The same results were obtained by TLD dosimetry on patient skin. Conclusion: Use of melt shielding methods is a very easy and fast shield-making technique with no differences in PDD, absolute and skin dose between lead and Cerrobend block shielding methods. é 2015, Shiraz University of Medical Sciences. All right reserved.
dc.language.isoEnglish
dc.relation.ispartofJournal of Biomedical Physics and Engineering
dc.subjectcobalt 60
dc.subjectArticle
dc.subjectcerrobend block
dc.subjectclinical article
dc.subjectcomparative effectiveness
dc.subjectcontrolled study
dc.subjectgeneral medical device
dc.subjecthead and neck cancer
dc.subjecthuman
dc.subjectionization chamber
dc.subjectparameters
dc.subjectphoton therapy
dc.subjectradiation depth dose
dc.subjectradiation dose distribution
dc.subjectradiology phantom
dc.subjectshielding method
dc.subjectskin dose
dc.subjectskin sparing effect
dc.subjectstandard lead block
dc.subjectthermoluminescence dosimetry
dc.titleComparison between clinically used irregular fields shielded by cerrobend and standard lead blocks