Show simple item record

dc.contributor.authorMesbahi, A
dc.contributor.authorReilly, AJ
dc.contributor.authorThwaites, DI
dc.date.accessioned2018-08-26T06:34:10Z
dc.date.available2018-08-26T06:34:10Z
dc.date.issued2006
dc.identifier.urihttp://dspace.tbzmed.ac.ir:8080/xmlui/handle/123456789/43922
dc.description.abstractMonte Carlo modeling of a linear accelerator is the first and most important step in Monte Carlo dose calculations in radiotherapy. We developed a photon beam model for Varian 2100EX for dose calculation purposes using MCNP4C Monte Carlo code. Results of our modeling were in close agreement with our measurements. The effect of beam width on percentage depth doses and beam profiles was studied. Our results showed that electron beam width could be tuned using large field beam profile at the depth of maximum dose.
dc.language.isoEnglish
dc.relation.ispartofApplied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
dc.subjectMonte Carlo Method
dc.subjectParticle Accelerators
dc.subjectPhotons
dc.subjectRadiotherapy Dosage
dc.subjectRadiotherapy Planning, Computer-Assisted
dc.titleDevelopment and commissioning of a Monte Carlo photon beam model for Varian Clinac 2100EX linear accelerator.
dc.typearticle
dc.citation.volume64
dc.citation.issue6
dc.citation.spage656
dc.citation.epage62
dc.citation.indexPubmed
dc.identifier.DOIhttps://doi.org/10.1016/j.apradiso.2005.12.012


Files in this item

FilesSizeFormatView

There are no files associated with this item.

This item appears in the following Collection(s)

Show simple item record